Test Frames
Specifications
MMx Series Material Testing Frames Model Number
MMS-500
MMS-1000
MMS-2500
MMS-5000
MMD-10K
MMD-30K
MMD-50K
N
500
1000
2500
5000
10,000
30,000
50,000
Load Capacity
kgf
50
100 225
250 562
500
1000 2250 0.001
3000 6750 0.001
5000
lbf
112
1124 0.001
11,250
mm/min 0.001 in/min 0.00004
0.001
0.001
0.001
Minimum Speed
0.00004
0.00004
0.00004
0.00004
0.00004
0.00004
mm/min 1525
1525
1525
1525
1525
1525
752
Maximum Speed
in/min 60
60
60
60
60
60
30
µm 0.250
0.250
0.250
0.250
0.05
0.025
0.025
Position Control Resolution
µin
9.8
9.8 953 37.5 762
9.8
9.8
1.9
0.9
0.9
mm 559
1257 49.5 1016
1257 49.5 1016
1270
1245
1220
Vertical Test Space 1
in
22
50
49
48
mm 381
1162 45.75
1137 44.75
1111 43.75
Total Crosshead Travel
in
15
30
40
40
mm 100
100
100
100
424 16.7
424 16.7
424 16.7
Throat
in
4
4
4
4
Accuracy Load Measurement Accuracy Position Measurement 2 Accuracy Strain Measurement Accuracy Crosshead Speed
Load Cell Sensor Dependent
±0.0002 inch (±5 μm)
±0.5% of reading down to 1/50 of full scale with ASTM E83 class B or ISO 9513 class 0.5 extensometer
±0.1% of set speed
Data Sampling
Hz
1 to 2000
Digital I/O
8 channels @ 1-5V
Extensometer Connections Analogue Inputs Analogue Outputs Electrical Phase
2 channels for 0-10V extensometers
1 channel @ ±10V 2 channels @ 0-10V
1
100, 120, 220, 230, 240Vac 10%
Single Phase Voltage (Vac) ±10% 220-240V
Single Phase Voltage (Vac) ±10% 220-240V
Power Requirements
100, 120, 220, 230, 240VAC 10%; 47-63Hz Self-identifying
°C °F °C °F
+10° to +38°C +50° to 100°F -40° to +66°C -40° to 150°F
Operating Temperature
Storage Temperature
Humidity
+10% to +90%, non-condensing
mm 805
1218 47.9
1573 61.9
1573 61.9
1685 66.4
1711 67.4
1711 67.4
Total Height
in
31.7
mm 381
381
381
381
787
787
787
Total Width
in
15
15
15
15
31
31
31
mm 514
514 20.3
514 20.3
514 20.3
724 28.5 136 300
724 28.5 192 425
724 28.5 225 500
Total Depth
in
20.3
kg
61
77
88
88
Weight
lb
135
170
195
195
NOTES Total vertical space is the distance from the top surface of the base plate to the bottom surface of the crosshead, excluding load cell sensor, test fixtures, and clevis adapter. Assumes Linear Error Correction and Deflection Compensation has been performed on test frame.
MMS and MMD test frames may be used with extensometers from Reliant Technologies and Epsilon Technology Corporation. Extensometers can be "plug & play" when supplied by the L.S. Starrett Company.
30